Email

AFP Rapid Quantitative Test Kit(CLIA)

The performance data

Manufacturer

SEKBIO

Parameter

AFP

Linearity

0.5-1000 ng/ml

Detection limit

0.5 ng/ml

benchmark product

Roche

Test principle and procedure

AFP: Add 10 μl of sample(The sample is diluted 10 times on the machine) + 30 μl of Ra+ 30 μl of Rd incubation for 15 minutes. Wash with 200 μl of washing solution. Finally, add 100 μl of A + 100 μl of B.


Contact Us Others

Six point calibration test results

Sample Concentration(ng/ml)

 

RLU

Average

CV

S/N

1162.62

4,132,851

4,047,944

4,090,398

1.47%

9130.35

644.64

2,413,626

2,402,215

2,407,921

0.34%

5374.82

137.78

552,114

578,005

565,060

3.24%

1261.29

13.00

56,977

59,955

58,466

3.60%

130.50

1.24

6,176

6,306

6,241

1.47%

13.93

0.04

453

443

448

1.58%

1.00

 

Precision

Sample

RLU

Concentration

Sample

RLU

Concentration

 

 

 

 

 

13

62,373

13.90

 

 

 

 

 

644.64

2,418,618

647.81

63,370

14.13

2,467,418

662.12

59,980

13.35

2,579,205

695.05

65,255

14.57

2,384,657

637.88

64,991

14.51

2,548,509

685.99

63,906

14.26

2,625,042

708.62

60,879

13.56

2,441,908

654.64

66,323

14.82

2,520,405

677.70

60,793

13.54

2,560,671

689.58

62,603

13.95

2,453,104

657.92

AVE

196,623

99.27

AVE

2,499,954

671.73

SD

2110

0.49

SD

78118

22.99

CV

1.07%

0.49%

CV

3.12%

3.42%

 

Limit of detection

Sample

RLU

Concentration

 

 

 

 

Ag Diluent

472

508

0.004

0.011

365

512

0.00

0.012

364

273

0.00

0.000

382

392

0.00

0.000

442

250

0.00

0.000

301

319

0.00

0.000

280

484

0.00

0.007

249

446

0.00

0.000

406

358

0.00

0.000

377

490

0.00

0.008

AVE

916

0.09

SD

120.35

0.05

AVE+2*SD

0.19


Linear

Sample(ng/ml)

Dilution ratio

Concentration value

Average

1303.80

1.0

1239.37

1265.76

1274.24

1303.799135

1173.42

0.9

1146.01

1214.25

1094.15

1151.25

1043.04

0.8

1032.30

1129.58

1131.06

1097.45

912.66

0.7

959.27

1032.66

931.43

974.29

782.28

0.6

801.08

847.83

801.65

816.81

651.90

0.5

691.66

704.87

637.77

678.01

521.52

0.4

539.84

496.45

489.92

508.68

391.14

0.3

391.37

394.63

374.61

386.86

260.76

0.2

287.37

292.72

295.31

291.80

130.38

0.1

136.52

133.42

124.08

131.33

0.02

0

0.00

0.00

0.00

0.02

 

AFP_Rapid_Quantitative_Test_Kit-1.jpg


Method Comparison


AFP_Rapid_Quantitative_Test_Kit-2.jpg

 

Reagents Stability

Reagent

Sample(ng/ml)

Average

Deviation

2-8°

1000.00

4,619,209


200.00

941,520


40.00

174,592


8.00

41,845


1.00

6,160


0.00

657


37℃ 1d

1000.00

4,541,645

-1.68%

200.00

951,361

1.05%

40.00

164,523

-5.77%

8.00

39,680

-5.18%

1.00

5,750

-6.66%

0.00

781

18.87%

37℃ 4d

1000.00

4,672,586

1.16%

200.00

946,372

-0.52%

40.00

175,190

6.48%

8.00

40,495

-3.23%

1.00

5,731

-0.32%

0.00

769

-1.54%

37℃ 7d

1000.00

4,334,765

-6.16%

200.00

900,764

-4.33%

40.00

152,647

-12.57%

8.00

40,724

-2.68%

1.00

5,335

-13.39%

0.00

684

4.11%

 

Calibrator Stability

Reagent

Sample(ng/ml)

Average

Deviation

2-8℃

1000.00

4,668,604


200.00

877,755


40.00

184,066


8.00

43,028


1.00

7,215


0.00

590


37℃ 1d

1000.00

4,643,345

1.1%

200.00

825,354

-6.8%

40.00

196,334

5.4%

8.00

44,191

2.8%

1.00

6,896

17.1%

0.00

567

-10.3%

37℃ 4d

1000.00

4,580,588

0.5%

200.00

1,056,764

-12.3%

40.00

183,616

12.5%

8.00

37,479

5.6%

1.00

5,374

11.9%

0.00

563

-13.7%

37℃ 7d

1000.00

3,882,314

-0.84%

200.00

824,447

12.24%

40.00

163,843

5.17%

8.00

37,187

-10.43%

1.00

5,881

-12.77%

0.00

593