Only 2%–5% of the cortisol in a blood sample is free, and the free fraction is the part that is biologically active. The rest is bound to plasma proteins. An antibody for a cortisol immunoassay has to work on a modest signal in serum, and on far less in saliva or hair.

Cortisol is the most abundant glucocorticoid secreted by the adrenal cortex. It helps regulate energy metabolism, electrolyte balance, blood pressure, immune responses and the stress response. Measuring it in body fluids is how laboratories diagnose Cushing's syndrome (too much cortisol) and Addison's disease (too little), and how they follow treatment.

Sekbio supplies two mouse monoclonal anti-cortisol antibodies, McAb-56 and McAb-45, plus a Cor-BSA conjugate to pair with them. The numbers below come from the Sekbio cortisol product sheet: detection limits, steroid cross-reactivity and a comparison with a commercial analyzer.

Scientist reading a cortisol immunoassay calibration curve on a time-resolved fluorometer next to a microplate and cortisol tubes
Figure 1. A cortisol immunoassay is read from a calibration curve, and the antibody sets both its detection limit and its specificity.

Why cortisol needs a competitive assay

Cortisol is a steroid of about 362 Da. Two IgG antibodies cannot sit on a molecule that small at the same time, so the sandwich format used for proteins is off the table. Cortisol assays are competitive: cortisol from the sample competes with a fixed amount of labeled or immobilized cortisol for a limited number of antibody binding sites.

The signal runs backwards. More cortisol in the sample means less signal. Figure 2 below shows exactly that shape.

Three things follow for anyone building a kit:

What a cortisol result is used for

The product sheet names the clinical questions a cortisol result answers:

The other direction is messier. Serum cortisol also rises with stress responses, psychiatric disorders, obesity, diabetes, alcoholism and pregnancy, and those physiological elevations can interfere with a Cushing's diagnosis. An assay cannot fix that. It can avoid adding errors of its own.

The sample type changes with the question. The Endocrine Society's Cushing's syndrome guideline (Nieman et al., J Clin Endocrinol Metab 2008) endorses urine free cortisol, late-night salivary cortisol and dexamethasone suppression testing as first-line tests. The primary adrenal insufficiency guideline (Bornstein et al., 2016) names the short corticotropin test as the reference test and morning ACTH and cortisol as an initial screen.

Clinical questionSample typesWhat it asks of the antibody
Cushing's syndrome screeningLate-night saliva, 24-hour urine, serum after dexamethasoneDetect low free-cortisol levels in saliva and urine
Adrenal insufficiencyMorning serum or plasma, before and after corticotropinQuantify across a wide range
Long-term stress exposureHairDetect the lowest concentrations of the three

Saliva, free cortisol and hair: where the detection limit counts

Total serum cortisol is the easy case. Free cortisol, which the sheet puts at 2%–5% of the total, is harder. Hair holds less again.

According to the sheet, immunoassays built on McAb-56 detect cortisol below 30 pg/mL. Besides blood and urine, they can also be used for the much lower levels of free cortisol and hair cortisol.

Hair is the extreme case because it records accumulated exposure, not one moment. A 2017 meta-analysis of 66 studies (Stalder et al.) found hair cortisol 22% higher in people under stress, and 43% higher when the stress was still ongoing at the time of sampling. Concentrations in hair extracts are far below serum levels, so sensitivity is not optional.

A DELFIA Detection Limit Does Not Transfer to Other Platforms

The "below 30 pg/mL" figure was measured in Cor-DELFIA, a time-resolved fluorescence format. A chemiluminescence, ELISA or lateral flow build will give a different limit with the same antibody. Re-measure it on your platform and in your sample matrix before you set a claim.

McAb-56 and McAb-45 side by side

Both clones are mouse monoclonal antibodies, and the sheet reports both in a Cor-DELFIA format.

Parameter (Cor-DELFIA)Anti-Cor McAb-56Anti-Cor McAb-45
Limit of detection< 30 pg/mL< 100 pg/mL
Cross-reactivity: corticosterone< 1%< 0.3%
Cross-reactivity: cortisone< 2.5%< 6%
Cross-reactivity: prednisolone< 79%< 10%
Other similar steroids at their highest physiological concentrationsNo effect on cortisol measurementNo effect on cortisol measurement

McAb-56 has the lower detection limit. McAb-45 is tighter on prednisolone, with a ceiling of 10% against 79%.

The gap on the other two steroids is smaller and runs both ways. McAb-45 is tighter for corticosterone (below 0.3% against below 1%), while McAb-56 is tighter for cortisone (below 2.5% against below 6%).

Read "<" as an Upper Bound

Every figure in this table is a ceiling reported on the sheet, not a measured midpoint. A clone listed as "< 79%" for prednisolone may cross-react less than that, but you should plan for the ceiling until you test it yourself.

Prednisolone is the one that bites in practice. It is a synthetic glucocorticoid that patients take for inflammatory disease, and a 2014 study of a widely used cortisol immunoassay (Krasowski et al.) found that prednisolone and 6-methylprednisolone cross-react strongly enough to matter in patients taking them. A result that reads high because of the drug can look like cortisol excess.

Our reading of the sheet: start with McAb-56 for saliva, free cortisol and hair work, where the detection limit decides the outcome, and with McAb-45 for serum assays aimed at patients on prednisolone. That is a reading of the numbers, not a tested recommendation. Run both clones against your own interference panel.

What the two plots on the sheet show

Calibration curve: signal falls as cortisol rises

Figure 2 is the typical Cor-DELFIA calibration curve for McAb-56, run from 0 to 10 ng/mL with six calibrator points.

Typical Cor-DELFIA cortisol calibration curve using anti-cortisol McAb-56, fluorescence in CPS against cortisol in ng/mL
Figure 2. Typical calibration curve for Cor-DELFIA using McAb-56, from the Sekbio cortisol product sheet.

Values read from the plot are approximate:

The curve is steep at the low end and flat at the top. Low cortisol levels are well resolved. Above about 4 ng/mL, a small signal error becomes a large concentration error.

Agreement with Elecsys Cortisol II

Figure 3 compares Cor-DELFIA built on McAb-56 with the Roche Elecsys Cortisol II assay on 24 samples.

Method comparison of Cor-DELFIA with anti-cortisol McAb-56 against Elecsys Cortisol II, regression y = 0.9671x + 1.8781, R2 = 0.9788, n = 24
Figure 3. Cortisol measured by Elecsys Cortisol II and by Cor-DELFIA with McAb clone 56, from the Sekbio cortisol product sheet.

A slope of 0.9671 puts Cor-DELFIA about 3% below Elecsys in proportion. The intercept of 1.8781 ng/mL is small against a range that reaches 490 ng/mL. For one immunoassay set against another, that is close agreement.

Agreement is not accuracy. Both methods are immunoassays and both can be pushed by cross-reacting steroids, so a good fit shows that they rank samples alike. When accuracy is the question, isotope-dilution LC-MS/MS is the usual reference.

What the product sheet does not tell you

It is a two-page sheet, so it leaves things out. Ask for these before you plan around it:

What Sekbio can supply

Three items are listed for cortisol immunoassay development:

Cortisol listings sit on the Sekbio cortisol antibody page. The same competitive logic is worked through for progesterone antibody pair design for competitive CLIA and LFA, and our VB12-BSA conjugate coating optimization post is a worked example of tuning a hapten-BSA conjugate for a competitive assay. For neighboring steroids, see the progesterone antibody and testosterone antibody pages, or browse the full IVD antibody and antigen catalog.

Frequently Asked Questions

What is cortisol and why is it measured?

Cortisol is the most abundant glucocorticoid made by the adrenal cortex. It is measured to diagnose Cushing's syndrome (too much cortisol) and Addison's disease (too little), and to monitor treatment, for example with the dexamethasone suppression test or hormone replacement. Stress, psychiatric disorders, obesity, diabetes, alcoholism and pregnancy can also raise it.

Why is cortisol measured with a competitive immunoassay instead of a sandwich?

Cortisol is a steroid of about 362 Da, too small for two antibodies to bind it at the same time. In a competitive assay, cortisol in the sample competes with a fixed amount of labeled or immobilized cortisol for a limited number of antibody sites, so the signal falls as cortisol rises. That also makes the single antibody responsible for the assay's specificity.

What detection limit do Sekbio's anti-cortisol antibodies reach?

In a Cor-DELFIA format, the limit of detection is below 30 pg/mL for McAb-56 and below 100 pg/mL for McAb-45. These figures come from the Sekbio cortisol product sheet and were measured on a time-resolved fluorescence platform. Limits change with platform and sample matrix, so re-measure on your own system.

Which steroids cross-react with McAb-56 and McAb-45?

For McAb-56, cross-reactivity is below 1% for corticosterone, below 2.5% for cortisone and below 79% for prednisolone. For McAb-45 it is below 0.3%, below 6% and below 10%. Other structurally similar steroid hormones at their highest physiological concentrations did not affect cortisol measurement with either clone.

Does prednisolone interfere with cortisol immunoassays?

It can. Prednisolone is structurally close to cortisol, and a 2014 study of a widely used cortisol immunoassay found that prednisolone and 6-methylprednisolone cross-react strongly enough to matter in treated patients. The Sekbio sheet gives prednisolone cross-reactivity below 79% for McAb-56 and below 10% for McAb-45, so McAb-45 is the safer first choice where prednisolone use is common.

Can these antibodies be used for saliva, urine or hair cortisol?

The sheet says McAb-56 immunoassays can measure cortisol in blood and urine, and can also be used for the much lower levels of free cortisol and hair cortisol. It does not include recovery or precision data for each matrix, so verify those in your own protocol before you rely on a result.

How well does Cor-DELFIA with McAb-56 agree with Elecsys Cortisol II?

On 24 samples, the regression was y = 0.9671x + 1.8781 with R² = 0.9788, with Cor-DELFIA on the y-axis and Elecsys Cortisol II on the x-axis. That is close agreement between two immunoassays. The sheet does not state the sample type, so treat it as a method-comparison snapshot rather than a full validation.

Can Sekbio supply cortisol antibodies and conjugate for kit development?

Yes. Anti-Cor McAb-56, Anti-Cor McAb-45 and Cor-BSA are listed for cortisol immunoassay development. Tell us your platform, sample type and target range and we can suggest a clone and plan a feasibility run. See our immunoassay development platforms or request a quote.

Which Clone to Request First

McAb-56 is the clone to try when the detection limit decides the project: saliva, free cortisol, hair. McAb-45 is the clone to try when your sample population includes patients taking prednisolone. Both come with the same gaps, because the sheet carries no matrix, precision or recovery data.

The practical next step is a side-by-side run on your own interference panel. Send us your platform, sample type and target range, and we can help plan it. Start from the Sekbio products page or contact the team.

Related Articles