CLIA · AE Platform · Competitive Assay

TT3 CLIA Bulk Reagent

Total Triiodothyronine Competitive Immunoassay for Thyroid Function Assessment

CLIA (AE)Competitive AssayTT3 / T30–9.18 nmol/LThyroid
0–9.18
nmol/L Calibration Range
≤6.8%
Within-Run CV (n=10)
2-Step
Competitive Assay Format
9.14×
S/N at Upper Calibrator

Competitive CLIA for Total T3 — Full Thyroid Panel Support

The Sekbio TT3 CLIA Bulk Reagent enables quantitative measurement of Total Triiodothyronine (TT3 / Total T3) — the biologically active thyroid hormone critical for metabolic regulation, cardiovascular function, and neurodevelopment. TT3 measurement, alongside TT4, fT3, fT4, and TSH, forms the comprehensive thyroid function panel for diagnosing hyperthyroidism, hypothyroidism, and thyroid hormone binding abnormalities.

Uses a two-step competitive immunoassay format on the AE chemiluminescence platform (Ra magnetic bead + Rc biotin + Rd AE). Calibration range 0–9.18 nmol/L covers the full TT3 reference interval (1.2–2.7 nmol/L) through hyperthyroid elevations. Within-run CV ≤6.8% meets clinical precision requirements.

Competitive CLIAAE Platform ISO 13485OEM Bulk Reagent

Analytical Specifications

Assay FormatCLIA Competitive (Two-Step)
Target AnalyteTotal T3 (TT3)
Sample TypeSerum / Plasma
Calibration Range0 – 9.18 nmol/L
CV (High QC ~4.71 nmol/L)6.79%
CV (Low QC ~1.83 nmol/L)6.66%
ReactionStep 1: Rd+sample 10min; Step 2: Ra+Rc 5min

Calibration & Precision

Two-step competitive: Step 1 — 30µL Rd + 30µL sample, 10 min (no wash); Step 2 — 30µL Ra + 30µL Rc, 5 min, wash; read RLU.

1 Six-Point Calibration Curve

Inverse dose-response characteristic of competitive assay — RLU decreases with increasing TT3 concentration. S/N defined as highest calibrator (0 nmol/L blank) divided by lowest calibrator RLU.

Calibrator (nmol/L)RLU Run 1RLU Run 2RLU Run 3Avg RLURelative Signal
0 (Max signal)1,933,5051,974,9551,999,3011,969,2541.00×
0.431,210,1961,222,3321,193,7931,208,7740.61×
1.83552,028526,336530,569536,3110.27×
2.91351,150346,261359,481352,2970.18×
4.71305,696303,464308,233305,7980.16×
9.18 (Min signal)217,174213,542215,716215,4770.11×
Competitive assay: signal decreases with analyte concentration. S/N ratio (0/9.18 nmol/L) = 9.14.
2 Precision — Within-Run (n=10)
QC Level (nmol/L)Avg Conc (nmol/L)SDCV
1.83 (Low)1.730.1156.66%
4.71 (High)4.000.2726.79%
n=10 replicates per level. Acceptance criterion: CV ≤10%.

Applications

Hyperthyroidism Diagnosis

Elevated TT3 (>2.7 nmol/L) is the primary marker for T3 thyrotoxicosis — where T3 is elevated while T4 may be normal. TT3 assay essential for Graves' disease workup and toxic nodular goiter evaluation.

Thyroid Function Panel

TT3 paired with TT4, fT3, fT4, and TSH forms the comprehensive thyroid function panel. Competitive CLIA format enables same-platform, same-analyzer thyroid profile automation.

Non-Thyroidal Illness (NTI) Assessment

TT3 falls in non-thyroidal illness (euthyroid sick syndrome) — monitoring TT3 alongside reverse T3 and TSH helps distinguish primary thyroid disease from systemic illness effects on thyroid hormones.

CLIA Thyroid Panel OEM

TT3 bulk reagent (Ra + Rc + Rd) for CLIA analyzer integration. Competitive format for small molecules. Available with TT4, fT3, fT4 pairs for complete thyroid panel OEM supply.

Request TT3 CLIA Reagent Information

Technical datasheets, performance data, and OEM pricing available on request.